Amperometric biosensor for oxalate determination in urine using sequential injection analysis

采用顺序注射分析法测定尿液中草酸盐的安培生物传感器

阅读:6
作者:Jose A Rodriguez, Prisciliano Hernandez, Veronica Salazar, Yolanda Castrillejo, Enrique Barrado

Abstract

An amperometric flow biosensor for oxalate determination in urine samples after enzymatic reaction with oxalate oxidase immobilized on a modified magnetic solid is described. The solid was magnetically retained on the electrode surface of an electrode modified with Fe (III)-tris-(2-thiopyridone) borate placed into a sequential injection system preceding the amperometric detector. The variables involved in the system such as flow rate, aspired volumes (modified magnetic suspension and sample) and reaction coil length were evaluated using a Taguchi parameter design. Under optimal conditions, the calibration curve of oxalate was linear between 3.0-50.0 mg·L⁻¹, with a limit of detection of 1.0 mg·L⁻¹. The repeatability for a 30.0 mg·L⁻¹ oxalate solution was 0.7%. The method was validated by comparing the obtained results to those provided by the spectrophotometric method; no significant differences were observed.