Halo assessment in intraocular lenses through high dynamic range images

通过高动态范围图像评估人工晶状体的光晕效应

阅读:1

Abstract

PURPOSE: To analyze the halo formation of several intraocular lenses (IOLs) in the optical bench. SETTING: University of Murcia, Murcia, Spain. DESIGN: In vitro study. METHODS: Light from a green LED passed through a pinhole and was collimated. Each IOL was placed within a realistic model eye having a PMMA cornea with a physiological amount of spherical aberration and a 4.5 mm aperture. A digital camera sensor acted as the retina and a focus tunable lens was used to change the object's vergence (range ±4 diopters). Series of images were captured with different exposure times and fused to get a high dynamic range image. Performance was assessed by analyzing the corresponding halo brightness and size. The tested lenses, that included biconvex and inverted meniscus IOLs, were monofocals, extended depth-of-focus (EDOF), and diffractive trifocals. RESULTS: Monofocal lenses produced halos with a radius close to 0.4 degrees. The halo radii of the nondiffractive EDOF lenses ranged between 0.45 and 0.63 degrees, whereas diffractive lenses had radii ranging from 0.84 to 1.22 degrees. The halo was generally dimmer for the refractive lenses and brighter for the diffractive. The through-focus images show that the halo size was larger at any defocus position for the diffractive lenses than for the rest of the tested IOLs. CONCLUSIONS: The diffractive IOLs exhibited a characteristic halo structure. Performance of the inverted meniscus and other nondiffractive lenses (Vivity and Eyhance) was comparable with a monofocal lens. This on-bench test can serve as an indication of the potential impact of photic phenomena on patient satisfaction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。