Abstract
PURPOSE: To compare and analyze the interchangeability of three measuring systems, each based on a different technique, for central corneal thickness (CCT) analysis. METHODS: CCT measurements were measured using optical coherence tomography (OCT), non-contact specular microscopy (NCSM), and ultrasonic pachymetry (USP) in 60 eyes of 60 healthy patients with a mean age of 66.5±15.0 years and a mean spherical equivalent of 0.43±1.14 D. Analysis of variations in measurement concordance and correlation among the three different methods were performed. Comparison of CCT measurements were done using Bland-Altman plots (with bias and 95% confidence intervals), intraclass correlation coefficient (ICC), and paired t-student analysis. RESULTS: Mean CCT values were: 549.20±26.91μm for USP (range 503-618μm), 514.20±27.49μm for NCSM (range 456-586μm) and 542.80±25.56μm for OCT (range 486-605μm). CCT values obtained with NCMS were significantly lower than those obtained with OCT and USP methods. NCMS CCT value was 36.08±10.72μm lower than USP value (p<0.05), and NCMS CCT value was 7.88±8.86μm lower than OCT value (p<0.05). ICC between USP-NCSM pair was 0.488 and 0.909 between USP-OCT pair. CONCLUSION: OCT and UPS offered highly comparable results, whereas NCSM offered lower mean CCT values compared to the other two methods. Therefore, NCSM should not be considered a reliable method for measuring CCT and should rather be considered for assessing longitudinal changes in the same patient.