Radiographic Texture Reproducibility: The Impact of Different Materials, their Arrangement, and Focal Spot Size

射线照相纹理再现性:不同材料、其排列方式和焦点尺寸的影响

阅读:1

Abstract

BACKGROUND: Feature reproducibility is a critical issue in quantitative radiomic studies. The aim of this study is to assess how radiographic radiomic textures behave against changes in phantom materials, their arrangements, and focal spot size. METHOD: A phantom with detachable parts was made using wood, sponge, Plexiglas, and rubber. Each material had 1 cm thickness and was imaged for consecutive time. The phantom also was imaged by change in the arrangement of its materials. Imaging was done with two focal spot sizes including 0.6 and 1.2 mm. All images were acquired with a digital radiography machine. Several texture features were extracted from the same size region of interest in all images. To assess reproducibility, coefficient of variation (COV), intraclass correlation coefficient (ICC), and Bland-Altman tests were used. RESULTS: Results show that 59%, 50%, and 4.5% of all features are most reproducible (COV ≤5%) against change in focal spot size, material arrangements, and phantom's materials, respectively. Results on Bland-Altman analysis showed that there is just a nonreproducible feature against change in the focal spot size. On the ICC results, we observed that the ICCs for more features are >0.90 and there were few features with ICC lower than 0.90. CONCLUSION: We showed that radiomic textures are vulnerable against changes in materials, arrangement, and different focal spot sizes. These results suggest that a careful analysis of the effects of these parameters is essential before any radiomic clinical application.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。