Galactose-1-phosphate uridyltransferase dried blood spot quality control materials for newborn screening tests

新生儿筛查试验用半乳糖-1-磷酸尿苷酰转移酶干血斑质量控制材料

阅读:1

Abstract

OBJECTIVES: We aimed to prepare dried-blood-spot (DBS) quality control (QC) materials for galactose-1-phosphate uridyltransferase (GALT), to evaluate their stability during storage and use, and to evaluate their performance in five DBS GALT test methods. DESIGN AND METHODS: We prepared and characterized GALT-normal and GALT-deficient DBS materials and compared GALT activities in DBSs after predetermined storage intervals at controlled temperatures and humidities. External evaluators documented the suitability of the DBS QC materials for use in five GALT test methods. RESULTS: GALT activity losses from DBSs stored in low (<30%) humidity for 14 days at 45°C, 35 days at 37°C, 91 days at room temperature, 182 days at 4°C, and 367 days at -20°C were 54%, 53%, 52% 23%, and 7% respectively. In paired DBSs stored in high humidity (>50%) for identical intervals, losses were: 45°C-68%; 37°C-79%; room temperature-72%, and 4°C-63%. GALT activities in DBSs stored at 4°C were stable throughout 19 excursions to room temperature. Twenty-five of 26 external evaluators, using five different GALT test methods, classified the GALT-deficient DBSs as "outside normal limits". All evaluators classified the GALT-normal DBSs as "within normal limits". CONCLUSIONS: Most of the GALT activity loss from DBSs stored at elevated or room temperature was attributable to the effects of storage temperature. Most of the loss from DBSs stored at 4°C was attributable to the effects of elevated humidity. Loss from DBSs stored at -20°C was insignificant. The DBS materials were suitable for monitoring performance of all five GALT test methods.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。