Cardiac-induced motion of the pancreas and its effect on image quality of ultrahigh-resolution CT

心脏运动引起的胰腺运动及其对超高分辨率CT图像质量的影响

阅读:2

Abstract

Recent advancements in diagnostic CT detector technology have made it possible to resolve anatomical features smaller than 20 LP/cm, referred to as ultra-high-resolution (UHR) CT. Subtle biological motions that did not affect standard-resolution (SR) CT may not be neglected in UHR. This study aimed to quantify the cardiac-induced motion of the pancreas and simulate its impact on the image quality of UHR-CT. We measured the displacement of the head of the pancreas in three healthy volunteers using Displacement Encoding with Stimulated Echoes (DENSE) MRI. The results were used to simulate SR- and UHR-CT acquisitions affected by pancreatic motion.We found pancreatic displacement in the 0.24-1.59 mm range during one cardiac cycle across the subjects. The greatest displacement was observed in the anterior-posterior direction. The time to peak displacement varied across subjects. Both SR and UHR images showed reduced image quality, as measured by radial modulation transfer function, due to cardiac-induced motion, but the motion artifacts caused more severe degradation in UHR acquisitions. Our investigation of cardiac-induced pancreatic displacement reveals its potential to degrade both standard and UHR-CT scans. To fully utilize the improvement in spatial resolution offered by UHR-CT, the effects of cardiac-induced motion in the abdomen need to be understood and corrected.Relevance statement Advancements in CT detector technology have enhanced CT scanner spatial resolution to approximately 100 µm. Consequently, previously ignored biological motions such as the cardiac-induced motion of the pancreas now demand attention to fully utilize this improved resolution.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。