Algorithms for a closed-loop artificial pancreas: the case for proportional-integral-derivative control

闭环人工胰腺算法:比例-积分-微分控制的案例

阅读:1

Abstract

Closed-loop insulin delivery continues to be one of most promising strategies for achieving near-normal control of blood glucose levels in individuals with diabetes. Of the many components that need to work well for the artificial pancreas to be advanced into routine use, the algorithm used to calculate insulin delivery has received a substantial amount of attention. Most of that attention has focused on the relative merits of proportional-integral-derivative versus model-predictive control. A meta-analysis of the clinical data obtained in studies performed to date with these approaches is conducted here, with the objective of determining if there is a trend for one approach to be performing better than the other approach. Challenges associated with implementing each approach are reviewed with the objective of determining how these approaches might be improved. Results of the meta-analysis, which focused predominantly on the breakfast meal response, suggest that to date, the two approaches have performed similarly. However, uncontrolled variables among the various studies, and the possibility that future improvements could still be effected in either approach, limit the validity of this conclusion. It is suggested that a more detailed examination of the challenges associated with implementing each approach be conducted.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。