Evaluation of Bone Wax Coated Bipolar Coagulation Forceps: Performance and Safety Assessment

骨蜡涂层双极凝固钳的评价:性能和安全性评估

阅读:8
作者:Jichun Shi, Wei Wei, Zhen Wang, Haobin Ren, Chenguang Jia, Lixin Dong, Zhengwei Li, Jianjian Zhang, Yu Feng, Kaixin Huang, Xiang Li, Jincao Chen

Background

Improving the performance of bipolar coagulation forceps is crucial for safer and more accurate neurosurgery. In our department, we found that bone wax (BW) melted by thermal effect of bipolar electrocoagulation can achieve more efficient hemostasis and reduce the amount of BW in neurosurgical procedures associated with bleeding from emissary and diploic veins. Nevertheless, relevant studies are still lacking to verify our finding.

Conclusion

This study reveals that BW coated bipolar coagulation forceps can provide a convenient, cost-efficient, safer, and more efficient way for hemostasis. More research is needed to evaluate the electrocoagulation with BW in the long term and verify our finding in human beings.

Methods

Tissue adhesion was evaluated by comparing the wetting tension and the amount of protein adhered to the forceps tips after electrocoagulation. Thermal damage was assessed by analyzing the thermography and H&E staining of coagulated rat brain tissues. The hemostatic efficiency was reflected by the number of electrocoagulation until complete hemostasis and the condition of damaged common carotid arteries. The safety of BW coated forceps in electrocoagulation was assessed by evaluating the inflammation of coagulated rat primary motor cortex and the motor functions at the 7th day postoperatively.

Objective

The study objectives were to evaluate the performance and safety in electrocoagulation: (1) compare the performance of BW coated bipolar coagulation forceps and the conventional anti-stick forceps in vivo, and (2) assess the safety of electrocoagulation with BW coated bipolar coagulation forceps in rat primary motor cortex.

Results

Bone wax coated forceps had a significantly higher contact angle and adhered less coagulum. Thermography was acquired at 3 s, 6 W units in rat primary motor cortex in vivo. The highest temperature recorded during BW coated tips application was significantly lower than the uncoated. In addition, there was a relatively smaller tissue injury area produced by the BW coated forceps. Additionally, BW coated forceps improved the hemostatic efficiency and caused fewer injuries on the damaged arteries (3 s, 10 W units). More importantly, electrocoagulation with BW coated forceps led to no significant motor function impairments and less glial and microglia responses.