Fixed Beamline Optimization for Intensity Modulated Carbon-Ion Therapy

强度调制碳离子治疗的固定光束线优化

阅读:1

Abstract

A major obstacle for the adoption of heavy ion therapy is the cost and technical difficulties to construct and maintain a rotational gantry. Many heavy ion treatment facilities instead choose to construct fixed beamlines as a compromise, which we propose to mitigate with optimized treatment couch angle. We formulate the integrated beam orientation and scanning spot optimization problem as a quadratic cost function with a group sparsity regularization term. The optimization problem is efficiently solved using fast iterative shrinkage-thresholding algorithm (FISTA). To test the method, we created the fixed beamline plans with couch rotation (FBCR) and without couch rotation (FB) for intensity modulated carbon-ion therapy (IMCT) and compared with the ideal scenario where both the couch and gantry have 360 degrees of freedom (GCR). FB, FBCR, and GCR IMCT plans were compared for ten pancreas cases. The FBCR plans show comparable PTV coverage and OAR doses for each pancreas case. In conclusion, the dosimetric limitation of fixed beams in heavy ion radiotherapy may be largely mitigated with integrated beam orientation optimization of the couch rotation.