Determinants and clinical impact of pressure drift in manoscan anorectal high resolution manometry system

肛门直肠高分辨率测压系统中压力漂移的决定因素及其临床影响

阅读:1

Abstract

BACKGROUND: Pressure drift (PD), resulting from differences between room and body temperature, reduces the accuracy of pressure measurements with the Manoscan high resolution manometry (HRM) system. Our aims were to assess PD during anorectal HRM. METHODS: Defined as the residual pressure measured immediately after the catheter was removed, PD was calculated for each sensor and averaged across all 12 sensors in 454 anorectal consecutive studies recorded with 3 HRM catheters. The relationship between PD and study duration, number of prior uses of a catheter, and peak and average pressure exposure during a study were evaluated. The correction of PD with a software algorithm (thermal compensation) was evaluated in 76 studies where the most distal sensor was outside the body. KEY RESULTS: The PD varied among sensors and across catheters. The average PD (7.3 ± 0.2 mmHg) was significantly greater for newer catheters, during longer studies, or when sensors were exposed to higher pressures. Together, these factors explained 81% of the variance in overall PD. After thermal compensation, the uncorrected median PD for the most distal sensor was 2.5-5 mmHg over the study duration. Correcting this changed the interpretation (e.g., as abnormal instead of normal) of at least 1 anorectal parameter in eight of 76 studies. CONCLUSIONS & INFERENCES: During anorectal HRM, PD declines with catheter use and is greater for newer catheters, when sensors are exposed to higher pressures, and for studies of longer duration. While PD is partially corrected with thermal compensation algorithms, the impact on interpretation is modest.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。