Novel Solid-State Potentiometric Sensors Using Polyaniline (PANI) as A Solid-Contact Transducer for Flucarbazone Herbicide Assessment

新型固态电位传感器,以聚苯胺(PANI)为固态接触换能器,用于氟唑磺隆除草剂评估

阅读:1

Abstract

Novel potentiometric solid-contact ion-selective electrodes (SC/ISEs) based on molecularly imprinted polymers (MIPs) as sensory carriers (MIP/PANI/ISE) were prepared and characterized as potentiometric sensors for flucarbazone herbicide anion. However, aliquat S 336 was also studied as a charged carrier in the fabrication of Aliquat/PANI/ISEs for flucarbazone monitoring. The polyaniline (PANI) film was inserted between the ion-sensing membrane (ISM) and the electronic conductor glassy carbon substrate (GC). The sensors showed a noticeable response towards flucarbazone anions with slopes of -45.5 ± 1.3 (r(2) = 0.9998) and -56.3 ± 1.5 (r(2) = 0.9977) mV/decade over the range of 10(-2)-10(-5), 10(-2)-10(-4) M and detection limits of 5.8 × 10(-6) and 8.5 × 10(-6) M for MIP/PANI/ISE and Aliguat/PANI/ISE, respectively. The selectivity and long-term potential stability of all presented ISEs were investigated. The short-term potential and electrode capacitances were studied and evaluated using chronopotentiometry and electrochemical impedance spectrometry (EIS). The proposed ISEs were introduced for the direct measurement of flucarbazone herbicide in different soil samples sprayed with flucarbazone herbicide. The results agree well with the results obtained using the standard liquid chromatographic method (HPLC).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。