Alteration of Membrane Permeability of Bacteria and Yeast by High Frequency Alternating Current (HFAC)

高频交流电(HFAC)对细菌和酵母细胞膜通透性的改变

阅读:1

Abstract

AIMS: Endox((R)) Endodontic System (Endox) is used for endodontic treatment by a high frequency alternating current (HFAC). This device damaged the envelopes of spores and vegetative organisms. If the integrity of the envelope is compromised, the transit of compounds in the two directions is possible. This latter aspect was investigated here. METHODS: The instrument delivered a 60ms pulse at a frequency 300 kHz, and power 800 KV/m. DNA transfer was verified using Escherichia coli K-12 strain carrying a non conjugative plasmid pBP517 (gyrA(+)) as donor and a rifampicin and nalidixic acid resistant recipient. 0.2 ml of mixture of donor and recipient strains in saline was exposed to HFAC and plated on selective media. Uptake of antimicrobials and a delay in re-growth was assessed exposing the strains to HFAC. RESULTS: Plasmid transfer was detected under different experimental conditions. From 9 to 27 recombinants were found. Representative recombinants cured from plasmid showed the original phenotype. HFAC promoted the uptake of ineffective antibiotics, and induces a 1 h of delay in re-growth on the strains. CONCLUSIONS: Endox exhibited an effect on microrganisms which is reminiscent with that occuring in electroporation, but with a mode of action that saved materials and time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。