Realizing the AF4-UV-SAXS on-line coupling on protein and antibodies using high flux synchrotron radiation at the CoSAXS beamline, MAX IV

利用 MAX IV 型 CoSAXS 光束线上的高通量同步辐射,实现了蛋白质和抗体的 AF4-UV-SAXS 在线耦合分析。

阅读:1

Abstract

In this paper, we demonstrate the coupling of synchrotron small angle X-ray scattering (SAXS) to asymmetrical flow-field flow fractionation (AF4) for protein characterization. To the best of our knowledge, this is the first time AF4 is successfully coupled to a synchrotron for on-line measurements on proteins. This coupling has potentially high impact, as it opens the possibility to characterize individual constituents of sensitive and/or complex samples, not suited for separation using other techniques, and for low electron density samples where high X-ray flux is required, e.g., biomolecules and biologics. AF4 fractionates complex samples in native or close to native environment, with low shear forces and system surface area. Many orders of magnitude in size can be fractionated in one measurement, without having to reconfigure the experimental setup. We report AF4 fractionations with correlated UV and statistically adequate SAXS data of bovine serum albumin and a monoclonal antibody and evaluate SAXS data recorded for the two protein systems.