Abstract

BACKGROUND AND PURPOSE: The published results of treating internal carotid artery aneurysms with the PED do not necessarily apply to its use in the posterior circulation because disabling brain stem infarcts can be caused by occlusion of a single perforator. In this multicenter study, we assessed the safety of PED placement in the posterior circulation. MATERIALS AND METHODS: A prospective case registry was maintained of all posterior circulation aneurysms treated with PEDs at 3 Australian neurointerventional centers during a 27-month period. The objective was to assess the complications and aneurysm occlusion rates associated with posterior circulation PEDs. RESULTS: Thirty-two posterior circulation aneurysms were treated in 32 patients. No deaths or poor neurologic outcomes occurred. Perforator territory infarctions occurred in 3 (14%) of the 21 patients with basilar artery aneurysms, and in all 3, a single PED was used. Two asymptomatic intracranial hematomas were recorded. No aneurysm rupture or PED thrombosis was encountered. The overall rate of permanent neurologic complications was 9.4% (3/32); all 3 patients had very mild residual symptoms and a good clinical outcome. Aneurysm occlusion was demonstrated in 85% of patients with >6 months of follow-up and 96% of patients with >1 year of follow-up. CONCLUSIONS: The PED is effective in the treatment of posterior circulation aneurysms that are otherwise difficult or impossible to treat with standard endovascular or surgical techniques, and its safety is similar to that of stent-assisted coiling techniques. A higher clinical perforator infarction rate may be associated with basilar artery PEDs relative to the internal carotid artery.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。