Dialysate flow rate and delivered Kt/Vurea for dialyzers with enhanced dialysate flow distribution

透析液流速和透析器输送的Kt/Vurea值(透析液流量分布增强型透析器)

阅读:1

Abstract

BACKGROUND AND OBJECTIVES: Previous in vitro and clinical studies showed that the urea mass transfer-area coefficient (K(o)A) increased with increasing dialysate flow rate. This observation led to increased dialysate flow rates in an attempt to maximize the delivered dose of dialysis (Kt/V(urea)). Recently, we showed that urea K(o)A was independent of dialysate flow rate in the range 500 to 800 ml/min for dialyzers incorporating features to enhance dialysate flow distribution, suggesting that increasing the dialysate flow rate with such dialyzers would not significantly increase delivered Kt/V(urea). DESIGN, SETTING, PARTICIPANTS, & MEASUREMENTS: We performed a multi-center randomized clinical trial to compare delivered Kt/V(urea) at dialysate flow rates of 600 and 800 ml/min in 42 patients. All other aspects of the dialysis prescription, including treatment time, blood flow rate, and dialyzer, were kept constant for a given patient. Delivered single-pool and equilibrated Kt/V(urea) were calculated from pre- and postdialysis plasma urea concentrations, and ionic Kt/V was determined from serial measurements of ionic dialysance made throughout each treatment. RESULTS: Delivered Kt/V(urea) differed between centers; however, the difference in Kt/V(urea) between dialysate flow rates of 800 and 600 ml/min was NS by any measure (95% confidence intervals of -0.064 to 0.024 for single-pool Kt/V(urea), -0.051 to 0.023 for equilibrated Kt/V(urea), and -0.029 to 0.099 for ionic Kt/V). CONCLUSIONS: These data suggest that increasing the dialysate flow rate beyond 600 ml/min for these dialyzers offers no benefit in terms of delivered Kt/V(urea).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。