Investigation of an optimized scanning protocol for the dentomaxillofacial region using 320-slice multidetector computed tomography

利用320层多层螺旋CT对口腔颌面部进行优化扫描方案的研究

阅读:1

Abstract

OBJECTIVES: To propose an imaging protocol that provides satisfactory image quality for oral examination while minimizing radiation dosage using 320-slice multidetector CT (MDCT). METHODS: An anthropomorphic head phantom was scanned using 320 MDCT with protocols combining different scanning modes: volume scanning (whole or local) vs helical scanning (80- or 64-slice detectors); tube voltage settings (80 kVp, 120 kVp and 135 kVp); and tube current settings (60 mA, 80 mA, 100 mA and 120 mA). A total of six anatomical bone structures and three anatomical soft-tissue structures were assessed using quantitative and qualitative analysis in the three orthographic planes (axial, sagittal and coronal). A figure of merit (FOM) was used to determine the optimal imaging protocol in terms of tube voltage, tube current and scanning mode. RESULTS: The 80-kVp setting had the worst quantitative and qualitative results (both p < 0.001) compared with the 135-kVp and 120-kVp settings, especially for soft-tissue structures. A significant difference was noted for the scores obtained using a tube current between 120 mA and 60 mA by quantitative analysis, but not by qualitative analysis. Volume scans using either whole or local modes had a significantly higher FOM than helical scanning of 80 or 64 slices. CONCLUSIONS: In 320 MDCT, a protocol using 135 kVp, 80 mA and the volume-scanning mode (whole or local) offers adequate visualization of both soft-tissue and bone structures while keeping the radiation dose as low as possible. This may therefore be considered a first choice among a wide selection of scanning protocols for dentomaxillofacial CT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。