Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study

采用电子后准直的狭缝扫描数字胸部X线摄影系统的散射抑制和低对比度性能:胸部模型研究

阅读:2

Abstract

Anti-scatter grids have been widely used to reject scatter and increase the perceptibility of low-contrast object in chest radiography; however they also attenuate the primary x-rays, resulting in a substantial degradation of primary information. Compensation for this degradation requires the use of higher exposure technique hence higher dose to the patient. A more efficient approach to reject scatter is the slot-scan imaging technique which employs a narrow scanning x-ray fan beam in conjunction with a slit or slot shaped solid state detector or an area detector used with an aft-collimator. With this approach, scatter can be rejected effectively without the need to attenuate primary x-rays. This paper demonstrates an electronic aft-collimation method, referred to as the alternate line erasure and readout (ALER) technique, for implementing the slot-scan digital radiography with a modern flat-panel detector. With this technique, instead of first exposing the detector and then reading the image line by line, the image line on the leading edge of the scanning fan beam is reset to erase the scatter accumulated prior to the arrival of the fan beam x-rays, while the image line on the trailing edge of the scanning fan beam is read out to acquire the image signals following the fan-beam exposure. These reset and readout processes are alternated and repeated as the x-ray fan beam scans across the detector. An anthropomorphic chest phantom was imaged to evaluate the scatter rejection ability and the low-contrast performance for the ALER technique and compare them with those for the anti-scatter grid method in full-field chest imaging. With a projected beam width of 16 mm, the slot-scan/ALER technique resulted in an average reduction of the scatter-to-primary ratios by 81%, 84%, 82%, and 86% versus 65%, 73%, 74%, and 73% with the anti-scatter grid method in the lungs, mediastinum, retrocardium, and subdiaphragm, respectively. The average CNR for the slot-scan/ALER technique was found to improve by 135%, 133%, 176%, and 87% versus 15%, 15%, 38%, and -11% with the anti-scatter grid method in the mediastinum, retrocardium, subdiaphragm, and lungs, respectively. These results demonstrated that the slot-scan/ALER technique can be used to achieve equally effective scatter rejection but substantially higher low-contrast performance than the anti-scatter grid method.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。