Multimethod Assessment of the Cyclic Fatigue Strength of ProGlider, Edge Glide Path and R-Pilot Endodontic Instruments

采用多种方法评估ProGlider、Edge Glide Path和R-Pilot根管治疗器械的循环疲劳强度

阅读:1

Abstract

BACKGROUND: The aim of this study was to comprehensively evaluate the cyclic fatigue strength of ProGlider, Edge Glide Path, and R-Pilot instruments. METHODS: Sixty-three instruments were submitted to a multimethod evaluation. Their design was analyzed by stereomicroscopy and scanning electron microscopy, including the number of blades, helical angle means, cross-sectional design, surface finishing, and symmetry. Energy-dispersive X-ray spectroscopy was used determine the nickel/titanium elements ratio, and differential scanning calorimetry determined the instruments' phase transformation temperatures. The cyclic fatigue tests were conducted in an artificial canal with a 6 mm radius and 86 degrees of curvature. The Mood's median test and one-way ANOVA were used to determine differences, with the significance level set at 0.05. RESULTS: The ProGlider presented the highest number of blades (n = 21), while R-Pilot had the highest helical angles (26.4°). Differences were noted in the instruments' cross-sections and surface finishing. The ProGlider and R-Pilot showed some similarities regarding the phase transformation temperatures but differed from the Edge Glide Path. All alloys showed an almost equiatomic nickel/titanium ratio. The R-Pilot instruments showed a significantly higher (p < 0.05) time to fracture than both the other files. CONCLUSION: Reciprocating R-Pilot instruments showed a higher cyclic fatigue time to fracture than the ProGlider and Edge Glide Path rotary files.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。