Validation and Application of an HPLC-UV Method for Routine Therapeutic Drug Monitoring of Cefiderocol

高效液相色谱-紫外检测法在头孢地洛常规治疗药物监测中的验证与应用

阅读:1

Abstract

Cefiderocol is a new siderophore cephalosporin approved for the treatment of multidrug resistant bacteria including activity against carbapenem-resistant Enterobacterales and Pseudomonas aeruginosa. As cephalosporins are known for their high pharmacokinetic variability in critically ill patients, cefiderocol therapeutic drug monitoring might become a valuable tool. Therefore, we aimed to develop and validate a simple, rapid, cost-effective high performance liquid chromatography (HPLC) method for the quantification of cefiderocol in serum. Samples were treated for protein precipitation followed by chromatographic separation on a reverse phase column (HPLC C-18) with gradient elution of the mobile phase. Cefiderocol was detected via UV absorption and quantification was performed with the internal standard (metronidazole) method. The calibration range showed linearity from 4 to 160 mg/L. The intra and interday precision was less than 10% with a recovery rate of 81%. The method was successfully used for the analysis of subsequent serum samples of critically ill patients and showed good performance in monitoring serum levels and optimizing antibiotic therapy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。