Developed wearable miniature sensor to diagnose initial perturbations of cardiorespiratory system

开发了一种可穿戴微型传感器,用于诊断心肺系统的早期扰动。

阅读:1

Abstract

The progress of microelectromechanical systems tends to fabricate miniature motion sensors that can be used for various purposes of biomedical systems, particularly on-body applications. A miniature wireless sensor is developed that not only monitors heartbeat and respiration rate based on chest movements but also identifies initial problems in the cardiorespiratory system, presenting a healthy measure defined based on height and length of the normal distribution of respiration rate and heartbeat. The obtained results of various tests are compared with two commercial sensors consisting of electrocardiogram sensor as well as belt sensor of respiration rate as a reference (gold standard), showing that the root-mean-square errors obtain <2.27 beats/min for a heartbeat and 0.93 breaths/min for respiration rate. In addition, the standard deviation of the errors reaches <1.26 and 0.63 for heartbeat and respiration rates, separately. According to the outcome results, the sensor can be considered an appropriate candidate for in-home health monitoring, particularly early detection of cardiovascular system problems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。