Applying the N-isopropylacrylamide gel dosimeter to quantify dynamic dose effects: A feasibility study

应用N-异丙基丙烯酰胺凝胶剂量计量化动态剂量效应:可行性研究

阅读:1

Abstract

BACKGROUND: The gel dosimeter is a chemical as well as a relative dosimeter. OBJECTIVE: To evaluate the feasibility of using N-isopropylacrylamide (NIPAM) gel dosimeter to observe the dynamic dose effects and quantification of the respiration, and to help determine the safety margins. METHODS: The NIPAM gel dosimeter combined with the dynamic phantom was used to simulate radiotherapy of lung or upper abdominal tumor. The field set to 4 × 5 cm2, simulate respiratory rate of 4 sec/cycle, and motion range 2 cm. MRI was used for reading, and MATLAB was used for analysis. The 3%/3 mm gamma passing rate > 95% was used as a clinical basis for evaluation. RESULTS: The dynamic dose curve was compared with 4 × 5, 4 × 4, 4 × 3 cm2 TPS, and gamma passing rates were 74.32%, 54.83%, 30.18%. Gamma mapping demonstrated that the highest dose region was similar to the result of the 4 × 4 cm2 TPS. After appropriate selection and comparing that the ⩾ 60% part of the dose curve with TPS, the gamma passing rate was 96.49%. CONCLUSIONS: Using the NIPAM gel dosimeter with dynamic phantom to simulate organ motion during respiration for dynamic dose measurement and quantified the dynamic dose effect is feasible. The results are consistent with clinical evaluation standards.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。