An extracellular matrix-mimetic coating with dual bionics for cardiovascular stents

一种用于心血管支架的具有双重仿生学的细胞外基质模拟涂层

阅读:1

Abstract

Anti-inflammation and anti-coagulation are the primary requirements for cardiovascular stents and also the widely accepted trajectory for multi-functional modification. In this work, we proposed an extracellular matrix (ECM)-mimetic coating for cardiovascular stents with the amplified functionalization of recombinant humanized collagen type III (rhCOL III), where the biomimetics were driven by structure mimicry and component/function mimicry. Briefly, the structure-mimic was constructed by the formation of a nanofiber (NF) structure via the polymerization of polysiloxane with a further introduction of amine groups as the nanofibrous layer. The fiber network could function as a three-dimensional reservoir to support the amplified immobilization of rhCoL III. The rhCOL III was tailored for anti-coagulant, anti-inflammatory and endothelialization promotion properties, which endows the ECM-mimetic coating with desired surface functionalities. Stent implantation in the abdominal aorta of rabbits was conducted to validate the in vivo re-endothelialization of the ECM-mimetic coating. The mild inflammatory responses, anti-thrombotic property, promotion of endothelialization and suppression of excessive neointimal hyperplasia confirmed that the ECM-mimetic coating provided a promising approach for the modification of vascular implants.