Perceptual Design and Evaluation of a Forearm-Based Vibrotactile Interface for Transfemoral Prosthetic Feedback

基于前臂的振动触觉界面在股骨假肢反馈中的感知设计与评估

阅读:1

Abstract

The lack of reliable sensory input from prosthetic limbs limits transfemoral amputees' ability to perceive limb movement without visual monitoring. This study evaluated design parameters of a proposed forearm-based vibrotactile system in a pre-clinical, design-level perceptual evaluation, conveying prosthetic joint positions through patterned vibrations to provide non-invasive proprioceptive feedback. Healthy participants completed two experiments assessing detection of tactile cues from dual-actuator bands on the wrist and elbow representing assumed ankle and knee positions. The effects of temporal structuring (sequential vs. simultaneous stimulation), actuator configuration, amplitude and frequency settings, and signal duration on response accuracy were examined. Sequential vibrations produced significantly higher recognition accuracy than simultaneous presentation (72.4% vs. 42.7%, p < 0.001) in a variety of vibration signal parameter values. Actuator placement also influenced performance: simultaneous stimulation on opposite forearm sides yielded significantly lower accuracy (p < 0.001) than same-side configurations, whereas this directional effect was not significant for sequential presentation. Accuracy did not differ significantly between equal and unequal amplitude or frequency levels across actuators. Longer stimulus durations improved accuracy, increasing from 82.3% at 60 ms to 92.5% at 240 ms, though the results indicated a saturation point, suggesting an optimal temporal window. These findings inform the design of forearm-based sensory feedback systems for improved prosthetic limb control.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。