Biocompatibility Evaluation of an Artificial Metallic Bone with Lattice Structure for Reconstruction of Bone Defect

用于骨缺损重建的具有晶格结构的金属人工骨的生物相容性评价

阅读:1

Abstract

Mandibular reconstruction for large bone defects is performed with consideration of patients' specific morphology and sufficient strength. Metal additive manufacturing techniques have been used to develop biomaterials for mandibular reconstruction. Titanium artificial mandibles with a lattice structure have been proposed, and the optimal conditions for their strength to withstand mechanical stress around the mandible have been reported. This study investigated the biocompatibility of a titanium artificial bone with a lattice structure fabricated under optimal conditions. The samples were fabricated using metal additive manufacturing. Body diagonals with nodes (BDN) were selected as suitable lattice structures. Dode medium (DM) was selected for comparison. The samples were implanted into rabbit tibial defects and resected with the surrounding bone at two and four weeks. Specimens were evaluated radiographically, histologically, and histomorphometrically. Radiopacity in each lattice structure was observed at two and four weeks. Histological evaluation showed trabecular bone-like tissue inside the BDN compared to the DM at four weeks. No significant differences were noted in the bone volume inside the structures. This study demonstrated the in vivo compatibility of artificial metallic bones with a BDN structure under mechanical stress conditions.