Large variations of oxygen delivery in self-inflating resuscitation bags used for preoxygenation - a mechanical simulation

用于预充氧的自充气复苏球囊氧气输送量存在较大差异——机械模拟

阅读:2

Abstract

BACKGROUND: Self-Inflating Resuscitation Bags (SIRB) are common and essential tools in airway management and ventilation. They are often used in resuscitation and emergency anaesthesia outside the operating theatre. There is a common notion that all SIRBs applied with a tight sealed mask will deliver close to 100 % oxygen during spontaneous breathing. The aim of the study was to measure the oxygen delivery of six commonly used SIRBs in a mechanical spontaneous breathing adult in vitro model. METHODS: Three SIRBs of each of the six models were evaluated for oxygen delivery during simulated breathing with an adult mechanical lung. The test was repeated three times per device (54 tests in total). The breathing profile was fixed to a minute volume of 10 L/min, a tidal volume of 500 mL and the SIRBs supplied with an oxygen fresh gas flow of 15 L/min. The fraction of delivered oxygen (FDO(2)) was measured over a three-minute period. Average FDO(2) was calculated and compared at 30, 60 and 90 s. RESULTS: At 90 s all models had reached a stable FDO(2). Average FDO(2) at 90 s; Ambu Oval Plus 99,5 %; Ambu Spur II 99,8 %; Intersurgical BVM Resuscitator 76,7 %; Laerdal Silicone 97,3 %; Laerdal The Bag II 94,5 % and the O-Two Smart Bag 39,0 %. All differences in FDO(2) were significant apart from the two Ambu models. CONCLUSIONS: In simulated spontaneous breathing, four out of six (by Ambu and Laerdal) Self-Inflating Resuscitation Bags delivered a high fraction of oxygen while two (Intersurgical and O-two) underperformed in oxygen delivery. These large variations confirm results reported in other studies. It is our opinion that underperforming Self-Inflating Resuscitation Bags might pose a serious threat to patients' health if used in resuscitation and anaesthesia. Manufacturers of Self-Inflating Resuscitation Bags rarely provide information on performance for spontaneous breathing. This poses a challenge to all organizations that need their devices to deliver adequate oxygen during spontaneous breathing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。