Customized-individually-made origin® implants in total knee arthroplasty allow a reliable solution for accurate reproduction of planned implant positioning

在全膝关节置换术中,定制的Origin®假体能够可靠地实现预定的假体植入位置。

阅读:1

Abstract

PURPOSE: To evaluate the accuracy and reproducibility of a patient-specific, customized individually made (CIM) total knee replacement (TKR) using the ORIGIN® prosthesis. METHODS: This was a prospective study conducted at a University Hospital from January 15, 2019, to April 30, 2021. The study included patients planned for an ORIGIN® CIM TKR procedure. Exclusion criteria included revision surgery, severe deformity, stiffness, or laxity. Evaluations were carried out using computed tomography scans performed 8 weeks preoperatively and 6 weeks postoperatively. The primary outcome measurements were the preoperative, planned, and postoperative CT scan alignment measurements including the Hip-Knee-Ankle (HKA) angle, mechanical Medial Distal Femoral articular surface Angle (mMDFA, distal alpha angle), Posterior Distal femoral articular surface angle (PDFA, posterior alpha angle), mechanical Medial Proximal Tibial articular surface Angle (mMPTA, beta angle) and posterior proximal tibial angle (PPTA). Secondary outcomes included the accuracy of implant positioning with percentage of outliers at 2° and 3° RESULTS: The study encompassed 51 knees from 50 patients with mean age of 68.1 (SD = 8.89). The overall HKA angle deviated by -0.93° [95% CI: -1.45; -0.43], and the PDFA angle by -0.61° [95% CI: -1.07; -0.15], while the mMPTA exceeded planned values by 1.00° [95% CI: 0.57; 1.43]. The 3° outliers rate ranged from 3.9% for the mMPTA to 7.8% for the HKA alignment, with no outliers in mMDFA and PPTA. Similarly, the 2° outliers rate ranged from 15.7% for both the PDFA angle and mMPTA to 19.6% for the HKA alignment. The Bland-Altman plots further emphasized the precision of planned and post-operative angles across all measurements. CONCLUSION: The CIM TKR showed high accuracy and reproducibility, closely matching preoperative planning. The weakest accuracy at 3°-outliers is in the reproduction of the HKA alignment at 92.2% (range for all angle: 92.2-100%). Similarly, the weakest accuracy at 2°-outliers is in the reproduction of the HKA alignment at 80.4% (range for all angles: 80.4-92.2%).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。