Stable hepatitis C virus RNA detection by RT-PCR during four days storage

在储存四天期间,通过RT-PCR检测丙型肝炎病毒RNA保持稳定。

阅读:1

Abstract

BACKGROUND: Suboptimal specimen processing and storage conditions of samples which contain hepatitis C virus (HCV) RNA may result in a decline of HCV RNA concentration or false-negative results in the detection of HCV RNA in serum. We evaluated the stability of HCV RNA in serum and clotted blood samples stored at room temperature or at 4 degrees C for 4 days with the aim of optimizing the standard procedures of processing and storage of samples. METHODS: Blood from five HCV RNA positive patients was collected in tubes with and without separator gel, centrifuged 1 or 6 hours after collection. Samples were then left 6, 24, 48, 72 or 96 h at room temperature (21.5 - 25.4 degrees C) or at 4 degrees C before determining their HCV RNA level using the COBAS AMPLICOR HCV MONITOR Test, vs 2.0 (Roche Diagnostic Systems). RESULTS: The logarithm of the HCV RNA level measurements remained within a 0.3 value of the means for 4 days at both temperatures (room temperature or 4 degrees C). CONCLUSIONS: We conclude that blood samples may be collected and aliquoted within 6 h of collection and can be stored at 4 degrees C for 72 hours as proposed by the manufacturer without significant differences in measured HCV RNA level. Our results indicate that lapses in this scheme may still yield reliable results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。