GlucoMen Day continuous glucose monitoring system: a screening for enzymatic and electrochemical interferents

GlucoMen Day 连续血糖监测系统:酶和电化学干扰物的筛查

阅读:11
作者:Fausto Lucarelli, Francesco Ricci, Felice Caprio, Francesco Valgimigli, Cosimo Scuffi, Danila Moscone, Giuseppe Palleschi

Background

While most of the common drugs with the potential to interfere with continuous glucose monitoring (CGM) systems are accessible over the counter and can be assumed by CGM patients without medical supervision, many other chemicals are frequently used to treat critically ill patients. Continuous glucose monitoring reading accuracy may also be compromised in patients characterized by abnormally high concentrations of physiological interferents. In this article, 22 species selected from endogenous and exogenous chemicals were screened as possible interferents of GlucoMen®Day (GMD), the new microdialysis-based CGM system from A. Menarini Diagnostics. Method: Interference testing was performed according to the EP7-A2 guideline (Clinical and Laboratory Standards Institute 2005). Interference was evaluated at two levels of glucose, with each interferent additionally tested at two concentrations. Furthermore, two configurations of the GMD disposable sensor kit--one designed for subcutaneous application, the other for direct intravascular CGM--were challenged with interferent-spiked serum and blood samples, respectively.

Conclusions

These results provide clear insights into the advantages offered by the use of a microdialysis-based CGM system that additionally relies on the detection of hydrogen peroxide at low operating potential. GlucoMen Day may become the CGM system of choice for those patients who require either regular administration of drugs or their glycemia to be tightly controlled in the intensive care unit or similar environments.

Results

With the exception of dopamine (however, at very high, nonphysiological concentrations), no interference was observed for all the tested substances. Interestingly, none of the common electrochemical interferents (including ascorbic acid, acetaminophen, and salicylic acid, which represent the major specificity issue for the competing CGM systems) significantly affected the system's output. Conclusions: These results provide clear insights into the advantages offered by the use of a microdialysis-based CGM system that additionally relies on the detection of hydrogen peroxide at low operating potential. GlucoMen Day may become the CGM system of choice for those patients who require either regular administration of drugs or their glycemia to be tightly controlled in the intensive care unit or similar environments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。