Conclusions
The QClamps® Plex UBA1 Mutation Detection Assay is a quicker, simpler, and more sensitive assay that can accurately detect the UBA1 mutations even at early stages with low mutation frequency.
Methods
Based on our proprietary XNA technology, we have developed a QClamp® Plex platform to detect eight mutations in a single reaction using the Luminex xMap technology. The assay sensitivity, specificity and precision were subsequently evaluated. Furthermore, the reference interval and clinical sensitivity/specificity were estimated using clinical healthy/positive DNA samples and the sanger sequencing method was used for comparison.
Results
With spiking synthetic mutant DNA in wildtype GM24385 cell line DNA, this assay can detect UBA1 mutations with a detection limit of variant allele frequency (VAF) at 0.1-5%. Our assay shows 100% concordance with Sanger sequencing results when used for analyzing 15 positive and 19 negative clinical samples. Conclusions: The QClamps® Plex UBA1 Mutation Detection Assay is a quicker, simpler, and more sensitive assay that can accurately detect the UBA1 mutations even at early stages with low mutation frequency.