Is there a "safe" suction pressure in the venous line of extracorporeal circulation system?

体外循环系统静脉管路中是否存在“安全”的抽吸压力?

阅读:1

Abstract

Successes of extracorporeal life support increased the use of centrifugal pumps. However, reports of hemolysis call for caution in using these pumps, especially in neonatology and in pediatric intensive care. Cavitation can be a cause of blood damage. The aim of our study was to obtain information about the cavitation conditions and to provide the safest operating range of centrifugal pumps. A series of tests were undertaken to determine the points at which pump performance decreases 3% and gas bubbles start to appear downstream of the pump. Two pumps were tested; pump R with a closed impeller and pump S with a semiopen impeller. The performance tests demonstrated that pump S has an optimal region narrower than pump R and it is shifted to the higher flows. When the pump performance started to decrease, the inlet pressure varies but close to -150 mmHg in the test with low gas content and higher than -100 mmHg in the tests with increased gas content. The same trend was observed at the points of development of massive gas emboli. Importantly, small packages of bubbles downstream of the pump were registered at relatively high inlet pressures. The gaseous cavitation in centrifugal pumps is a phenomenon that appears with decreasing inlet pump pressures. There are a few ways to increase inlet pump pressures: (1) positioning the pump as low as possible in relation to the patient; (2) selecting appropriate sized venous cannulas and their careful positioning; and (3) controlling patient's volume status.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。