Clinical comparison of BACTEC 9240 plus aerobic/F resin bottles and the isolator aerobic culture system for detection of bloodstream infections

BACTEC 9240 plus 需氧/F 树脂瓶与隔离器需氧培养系统在检测血流感染方面的临床比较

阅读:1

Abstract

The Plus Aerobic/F resin bottle of the BACTEC 9240 automated blood culture system (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) was compared with aerobic culture of the Isolator system (Wampole Laboratories, Cranbury, N.J.) for the detection of bloodstream microorganisms from 6,145 blood cultures collected from adult patients with suspected septicemia. The BACTEC resin bottles were incubated for 7 days, and the sediment from the Isolator tube was inoculated to sheep blood and chocolate agars which were incubated for 72 h and to inhibitory mold, brain heart infusion, and Sabouraud agars which were incubated for 21 days. A total of 622 microorganisms were recovered from 583 blood cultures. The BACTEC resin bottle recovered statistically significantly more pathogens overall than the Isolator system (P = 0.0006). When individual pathogens isolated from either system for a 7-day study period were assessed, it was determined that the BACTEC resin bottle detected statistically significantly more isolates of Staphylococcus aureus (P = 0.0113) and coagulase-negative Staphylococcus spp. (P = 0.0029) than the Isolator system. The BACTEC resin bottle also detected statistically significantly more bloodstream infections (septic episodes) caused by coagulase-negative Staphylococcus spp. (P = 0.0146). The Isolator system recovered statistically significantly more contaminants overall (P < 0.0001), and among this group of microorganisms, recovered statistically significantly more Bacillus spp. (P < 0.0001), coagulase-negative Staphylococcus spp. (P < 0.0001), and viridans group Streptococcus spp. (P = 0.0156). The Isolator system detected statistically significantly more isolates of Histoplasma capsulatum (P = 0.004), but all of these isolates were detected at > or = 7 days of incubation of fungal plates, i.e., after the system to system comparison study period (7 days). In blood culture sets which produced growth of the same pathogen in both systems, there was a statistically significant difference in median time to detection for all pathogens combined favoring the BACTEC resin bottle over the Isolator tube (P < 0.05). When assessing individual microorganisms, the median times for detection of S. aureus, Enterococcus spp., and Pseudomonas spp. were all statistically significantly less for the BACTEC system (P < 0.05). The BACTEC instrument had 79 (1.3%) false positive signals. The BACTEC system required less processing time than the Isolator system and eliminates the hands-on time for detection of positive cultures required with the Isolator system.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。