CSDR Coupling with Exo III for Ultrasensitive Electrochemistry Determination of miR-145

CSDR 与 Exo III 偶联用于 miR-145 的超灵敏电化学测定

阅读:10
作者:Moli Zhang, Yang Yang, Lingyi Xin, Hua Zhang, Lun Wu, Jun Zhu, Jing Zhu, Shiyun Liu, Zhaohui Wang, Qinhua Chen, Guangyi Yang

Abstract

Recently, miRNAs have become a promising biomarker for disease diagnostics. miRNA-145 is closely related to strokes. The accuracy determination of miRNA-145 (miR-145) in stroke patients still remains challenging due to its heterogeneity and low abundance, as well as the complexity of the blood matrix. In this work, we developed a novel electrochemical miRNA-145 biosensor via subtly coupling the cascade strand displacement reaction (CSDR), exonuclease III (Exo III), and magnetic nanoparticles (MNPs). The developed electrochemical biosensor can quantitatively detect miRNA-145 ranging from 1 × 102 to 1 × 106 aM with a detection limit as low down as 100 aM. This biosensor also exhibits excellent specificity to distinguish similar miRNA sequences even with single-base differences. It has been successfully applied to distinguish healthy people from stroke patients. The results of this biosensor are consistent with the results of the reverse transcription quantitative polymerase chain reaction (RT-qPCR). The proposed electrochemical biosensor has great potential applications for biomedical research on and clinical diagnosis of strokes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。