Electron probe x-ray analysis of single ferritin molecules

利用电子探针X射线分析单个铁蛋白分子

阅读:1

Abstract

Single molecules and groups of two or three ferritin molecules were subjected to electron probe x-ray microanalysis in a transmission electron microscope equipped with a liquid nitrogen cooled stage. Significant Fe Kalpha peaks were generated during 100-sec counts when single ferritin molecules were excited with a probe current of 0.35 nA/60 nm spot, less than the maximal current available in a thermionic gun. There was a linear relationship between the number of ferritin molecules analyzed and count rates. The experimental results are compared to the theoretically calculated Fe Kalpha yields and to the results of Isaacson and Johnson [(1975) Ultramicroscopy I, 33-52] with electron energy loss analysis. We conclude that current state of the art electron probe x-ray analysis can realize the theoretically predicted sensitivity of the method, and estimate 0.9 X 10(-19) g of Fe as the minimal mass detectable with maximal (thermionic) probe current during a 100-sec count and with 95% confidence.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。