Comparison of Arkray/ELITech ADAMS HA-8180V with Bio-Rad Variant, II Turbo2.0 and Tosoh Bioscience HLC-723G8 for HbA1c determination

比较 Arkray/ELITech ADAMS HA-8180V、Bio-Rad Variant II Turbo2.0 和 Tosoh Bioscience HLC-723G8 在 HbA1c 测定中的性能

阅读:1

Abstract

OBJECTIVE: This study aimed to evaluate a new ion-exchange HPLC system to measure hemoglobin A1c (HbA1c) in comparison to two other widely used HPLC systems. METHODS: Analytical performance and hemoglobin variants detection was assessed on the new Arkray/ELITech ADAMS HA-8180V, in parallel to Bio-Rad Variant II Turbo2.0 and Tosoh Bioscience HLC-723G8. A method comparison and concordance for the classification of patients according to the American Diabetes Association (ADA) categories was performed using kappa test. RESULTS: ADAMS HA-8180V demonstrated excellent within-run imprecision (0%) and between-run imprecision: ≤1.21% using blood sample and ≤0.94% using quality controls. Method comparison of ADAMS HA-8180V with the two other systems yielded very high coefficient correlation (r > 0.995). A very good agreement (kappa value ≥0.81) was observed between methods for the classification of patients according to the ADA categories. In addition, all systems were able to detect the presence of most classical variants such as HbC, HbS, HbD, and HbE. CONCLUSION: The Arkray/ELITech ADAMS HA-8180V demonstrated high analytical performance similar to previous systems such as Biorad(TM) Variant II Turbo 2.0 and Tosoh Bioscience HLC-723G8. The three systems allow a high-quality HbA1c measurement and appeared interchangeable for diabetes diagnosis or for the therapeutic monitoring of patients without hemoglobin variants.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。