Associated factors and outcomes of crossover from a laser sheath to a bidirectional rotational mechanical sheath during transvenous lead extraction

经静脉导线拔除术中从激光鞘管过渡到双向旋转机械鞘管的相关因素和结果

阅读:1

Abstract

BACKGROUND: During transvenous lead extraction (TLE), a GlideLight laser sheath (Philips) cannot always be advanced over the lead, and crossover to the Evolution system (i.e., an Evolution RL sheath or Evolution Shortie RL sheath [Cook Medical]) is required. We aimed to determine the associated factors and outcomes of such device crossover. METHODS: This observational study included 112 patients who underwent TLE. The patients were divided into crossover and non-crossover groups. Outcomes and associated factors of crossover were evaluated. RESULTS: Overall, 57 (50.9%) patients required crossover to the Evolution system (crossover group), whereas 55 (49.1%) patients did not require crossover (non-crossover group). Clinical success rate was similar between the two groups (98.3% vs. 100%; p = 1.00). No major intraprocedural complications related to powered sheaths occurred. Multivariate logistic regression analysis results showed that dwell time of the oldest extracted lead (per year) (odds ratio [OR]: 1.18, 95% confidence interval [CI]: 1.02-1.36; p = .026), number of leads extracted per procedure (OR: 7.23, 95% CI: 1.74-29.99; p = .007), and use of a femoral approach (OR: 21.09, 95% CI: 2.33-190.67; p = .007) were associated factors of crossover. The cutoff for crossover was 7.7 years from the implant (sensitivity 90.5%, specificity 64.9%, area under the curve 0.80). CONCLUSIONS: Both groups showed a high rate of clinical success. Switching to the Evolution system may facilitate a safe and effective TLE when a laser sheath does not advance despite laser activation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。