Comparative Stiffness Characteristics of Ilizarov- and Hexapod-type External Frame Constructs

伊利扎洛夫型和六足机器人型外框架结构的刚度特性比较

阅读:1

Abstract

BACKGROUND: The Ilizarov method and fixator are clinically recognised for the treatment of fractures, limb salvage and deformity correction. There have been extensive studies determining the basic mechanism for fracture healing using this technique. It is generally accepted that circular frames optimise the mechanical environment by reducing shear strain across the fracture while maintaining axial micromotion so as to promote fracture healing. There have been several new hexapod-type frames introduced into the market over the past 20 years with little comparative research into their biomechanical properties and resultant effects on the fracture environment. QUESTIONS/PURPOSES: To investigate the biomechanical behaviours of the TrueLok-Hex (TL-HEX) and Taylor spatial frame (TSF) hexapod-type circular external fixators with comparison to traditional Ilizarov-type (TL-Ilizarov and TSF-Ilizarov) constructs and potential performance in vivo. METHODS: Testing was performed on standardised four-ring TSF and TL-HEX constructs matched by identical frames using Ilizarov threaded rod constructs for each set of components. All frames were tested under physiological levels of axial, bending and torsional loading. Load-deformation properties for each construct under each mode of loading were calculated and analysed statistically using ANOVA. RESULTS: Under axial loading, the Ilizarov construct utilising TL-HEX components demonstrated the greatest rigidity followed by the Ilizarov construct using TSF components. Under bending loads, the difference in rigidity between constructs was similar but less marked. Under torsional loading, both hexapod frames were seen to be significantly more rigid than the Ilizarov constructs. Overall deformation around neutral loading was much higher in the TSF frame due to an observed significant "toe-in" laxity in the strut universal joints. The remaining deformation of both hexapod frames was similar with a higher level of TL-HEX rigidity in axial loading and a higher level of TSF rigidity in bending and torsion. CONCLUSION: In conclusion, both hexapod frame constructs were less rigid under axial loading but more rigid under bending and torsional loads than their comparative Ilizarov constructs. As a result of their Cardan universal joints, the TSF demonstrated greater overall planar strain due to the observed "toe-in" laxity around neutral loading while the TL-HEX, with ball-and-socket universal joints, demonstrated a minimal level of laxity. Beyond the initial deformation due to the preloaded laxity, both hexapod frames responded to loading in a similar manner. There were significant differences in the frames' mechanical behaviour under different loading conditions but further research is required to determine whether these translate in vivo into clinical significance. HOW TO CITE THIS ARTICLE: Fenton C, Henderson D, Samchukov M, et al. Comparative Stiffness Characteristics of Ilizarov- and Hexapod-type External Frame Constructs. Strategies Trauma Limb Reconstr 2021;16(3):138-143.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。