Real-time Monitoring of Contrast-enhanced Ultrasound for Radio Frequency Ablation

射频消融术对比增强超声的实时监测

阅读:1

Abstract

BACKGROUND: This study compared the realtime monitoring effects of conventional ultrasound and contrast-enhanced ultrasound (CEUS) on evaluating radio frequency ablation (RFA) in a living swine liver model. METHODOLOGY: Liver RFA was performed on 10 young swine. Conventional ultrasound and CEUS were performed immediately. After the animals were sacrificed, ablation lesions were removed to histopathologically examine the range of the lesions. Ablation completeness based on three methods were compared using histopathology as the gold standard. RESULTS: Forty-three ablation lesions were produced in the animals. The horizontal diameter, vertical diameter and ablation lesion area based on conventional ultrasound were all significantly smaller than those based on the gross sample, but no significant differences existed between the results of the CEUS and the gross sample. Histopathology showed that 30 lesions were incompletely ablated and 13 were completely ablated, while CEUS showed that 28 lesions were incompletely ablated and 15 were completely ablated. Compared with histopathology, CEUS had an accuracy of 81.4%, a sensitivity of 83.3%, and a specificity of 76.9%. No significant difference in ablation completeness judgment between CEUS and histopathology was observed. CONCLUSION: CEUS provides a real-time radiological foundation for evaluating RFA lesion ranges and completeness in a living swine liver model.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。