Effect of anatomical variation on extracorporeal membrane oxygenation circulatory support: A computational study

解剖变异对体外膜肺氧合循环支持的影响:一项计算研究

阅读:1

Abstract

BACKGROUND: Extracorporeal membrane oxygenation (ECMO) via femoral cannulation is a vital intervention capable of rapidly restoring perfusion for patients in shock. Despite increasing use to provide circulatory support, its hemodynamic effects are poorly understood and the impact of patient-specific anatomical variation on perfusion is unknown. This study investigates the complex failing heart-mechanical circulatory support circulation and analyzes the effect of patient-specific vascular anatomical variations on hemodynamics and end-organ perfusion. METHODS: Patient-specific vascular geometries were constructed from segmenting clinical computerized tomography angiography images and quantitatively compared using tortuosity, curvature, torsion, and lumen diameter. Computational fluid dynamic simulations were performed on a subset of geometries selected to represent a range of anatomical variation. Heart failure severity was modeled by varying the relative fraction of total flow provided by the heart and the extracorporeal circuit. A 3-element lumped parameter model was applied to accurately and dynamically model distal perfusion boundary conditions. Hemodynamic parameters and end-organ perfusion were analyzed and compared to assess the effect of anatomical variation. RESULTS: Pulsatile antegrade cardiac perfusion and ECMO retrograde perfusion collide in the aorta to form a dynamic watershed region. The size, position, and variation of this region over the cardiac cycle is substantially altered by patient anatomical region. Increased vascular tortuosity reduces the proximal extent of flow from circulatory support and decreases the size of the watershed region. CONCLUSIONS: Patient vascular anatomy is a key determinant of the ECMO-failing heart circulation that alters the location and extent of the watershed region and affects the tissues at risk for differential hypoxia and circuit-derived thromboemboli for a given level of support.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。