Deduction of the operable design space of RP-HPLC technique for the simultaneous estimation of metformin, pioglitazone, and glimepiride

反相高效液相色谱法同时测定二甲双胍、吡格列酮和格列美脲的可操作设计空间的推导

阅读:5
作者:Aya A Marie, Sherin F Hammad, Mohamed M Salim, Mahmoud M Elkhodary, Amira H Kamal

Abstract

A reversed-phase RP-HPLC method was developed for the simultaneous determination of metformin hydrochloride (MET), pioglitazone (PIO), and glimepiride (GLM) in their combined dosage forms and spiked human plasma. Quality risk management principles for determining the critical method parameters (CMPs) and fractional factorial design were made to screen CMPs and subsequently, the Box-Behnken design was employed. The analytical Quality by Design (AQbD) paradigm was used to establish the method operable design region (MODR) for the developed method depended on understanding the quality target product profile (QTPP), analytical target profile (ATP), and risk assessment for different factors that affect the method performance to develop an accurate, precise, cost-effective, and environmentally benign method. The separation was carried out using a mobile phase composed of methanol: 0.05 M potassium dihydrogen phosphate buffer pH 3.7 with 0.05% TEA (78:22, v/v). The flow rate was 1.2 mL/min. DAD detector was set at 227 nm. Linagliptin (LIN) was used as an internal standard. The proposed method was validated according to The International Council for Harmonization of Technical Requirements for Pharmaceuticals for Human Use (ICH). The assay results obtained by using the developed method were statistically compared to those obtained by the reported HPLC method, and a satisfying agreement was observed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。