Dosimetric impact of stopping power for human bone porosity with dual-energy computed tomography in scanned carbon-ion therapy treatment planning

双能计算机断层扫描在扫描碳离子治疗计划中对人体骨骼孔隙率的阻止本领剂量学影响

阅读:1

Abstract

Few reports have documented how the accuracy of stopping power ratio (SPR) prediction for porous bone tissue affects the dose distribution of scanned carbon-ion beam therapy. The estimated SPR based on single-energy computed tomography (SECT) and dual-energy CT (DECT) were compared for the femur of a Rando phantom which simulates the porosity of human bone, NEOBONE which is the hydroxyapatite synthetic bone substitute, and soft tissue samples. Dose differences between SECT and DECT were evaluated for a scanned carbon-ion therapy treatment plan for the Rando phantom. The difference in the water equivalent length was measured to extract the SPR of the examined samples. The differences for SPR estimated from the DECT-SPR conversion were small with - 1.8% and - 3.3% for the Rando phantom femur and NEOBONE, respectively, whereas the differences for SECT-SPR were between 7.6 and 70.7%, illustrating a 1.5-mm shift of the range and a dose difference of 13.3% at maximum point in the evaluation of the dose distribution. This study demonstrated that the DECT-SPR conversion method better estimated the SPR of the porosity of bone tissues than SECT-SPR followed by the accurate range of the carbon-ion beams on carbon-ion dose calculations.