D-shaped plastic optical fibre aptasensor for fast thrombin detection in nanomolar range

形塑料光纤适体传感器用于纳摩尔范围内快速检测凝血酶

阅读:5
作者:Nunzio Cennamo, Laura Pasquardini, Francesco Arcadio, Lia E Vanzetti, Alessandra Maria Bossi, Luigi Zeni

Abstract

The development of optical biosensors for the rapid and costless determination of clinical biomarkers is of paramount importance in medicine. Here we report a fast and low-cost biosensor based on a plasmonic D-shaped plastic optical fibre (POF) sensor derivatized with an aptamer specific for the recognition of thrombin, the target marker of blood homeostasis and coagulation cascade. In particular, we designed a functional interface based on a Self Assembled Monolayer (SAM) composed of short Poly Ethylene Glycol (PEG) chains and biotin-modified PEG thiol in ratio 8:2 mol:mol, these latter serving as baits for the binding of the aptamer through streptavidin-chemistry. The SAM was studied by X-ray Photoelectron Spectroscopy (XPS) analysis, static contact angle (CA), Surface Plasmon Resonance (SPR) in POFs, and fluorescence microscopy on gold surface. The optimized SAM composition enabled the immobilization of about 112 ng/cm2 of aptamer. The thrombin detection exploiting POF-Aptasensor occurred in short times (5-10 minutes), the reached Limit of Detection (LOD) was about 1 nM, and the detection range was 1.6-60 nM, indicating the POF-Aptasensor well addresses the needs for a low-cost, simple to use and to realize, rapid, small size and portable diagnostic platform.