Continuous, in situ detection of hydrogen sulfide (H(2)S) in biological milieu is made possible with electrochemical methods, but direct amperometry is constrained by the generation of elemental sulfur as an oxidative byproduct. Deposition of a sulfur layer passivates the working electrode, reducing sensitivity and causing performance variability. Herein, we report on the use of a surface preconditioning procedure to deposit elemental sulfur on a glassy carbon electrode prior to measurement and evaluate performance with common analytical metrics. The lack of traditional anti-poisoning techniques (e.g. redox mediators, cleaning pulses) also allowed for facile surface modification with electropolymerized films. For the first time, a series of electropolymerized films were characterized for their H(2)S permselective behavior against common biological interferents. Highly selective, film-modified electrodes were then evaluated for their anti-biofouling ability in simulated wound fluid. The final optimized electrode was capable of measuring H(2)S with a low detection limit (i.e., <100 nM) and ∼80% of its initial sensitivity in proteinaceous media.