Heterogeneous Immunoassay Using Channels and Droplets in a Digital Microfluidic Platform

利用数字微流控平台中的通道和液滴进行异质免疫分析

阅读:1

Abstract

This work presents a heterogeneous immunoassay using the integrated functionalities of a channel and droplets in a digital microfluidic (DMF) platform. Droplet functionality in DMF allows for the programmable manipulation of discrete sample and reagent droplets in the range of nanoliters. Pressure-driven channels become advantageous over droplets when sample must be washed, as the supernatant can be thoroughly removed in a convenient and rapid manner while the sample is immobilized. Herein, we demonstrate a magnetic bead-based, enzyme-linked immunosorbent assay (ELISA) using ~60 nL of human interleukin-6 (IL-6) sample. The wash buffer was introduced in the form of a wall-less virtual electrowetting channel by a syringe pump at the flow rate of 10 μL/min with ~100% bead retention rate. Critical parameters such as sample wash flow rate and bead retention rate were optimized for reliable assay results. A colorimetric readout was analyzed in the International Commission on Illumination (CIE) color space without the need for costly equipment. The concepts presented in this work are potentially applicable in rapid neonatal disease screening using a finger prick blood sample in a DMF platform.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。