Possibility of simultaneously measuring low and high calcium concentrations using Fura-2 and lifetime-based sensing

利用Fura-2和基于寿命的传感技术,可以同时测量低钙浓度和高钙浓度。

阅读:1

Abstract

We characterized the fluorescence probe Fura-2 for calcium measurements using frequency-domain phase-modulation fluorometry. By the use of different excitation wavelengths from 345 to 380 nm, the apparent calcium dissociation constants can be altered from 41 nM to 1.92 microM Ca2+. This change in apparent Kd results from changes in the relative extent of excitation of the calcium-bound and calcium-free forms, and the excitation wavelength-dependent contribution of each form to the intensity decay. These results indicate that lifetime-based measurements with Fura-2 can be used for imaging of calcium over a wide range of concentrations. An additional favorable feature of Fura-2 is that the calcium-free form can be almost exclusively excited at wavelength of 390 nm or longer, and can thus be used as a reference providing the lifetime in the absence of calcium, without removing the calcium. Additionally, exposure of Fura-2 to intense illumination shifts but does not distort the frequency response. For cellular imaging, these favorable properties of Fura-2 may allow calibration of the calcium concentrations without the use of ionophores.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。