Use of a sample injection loop for an accurate measurement of particle number concentration by flow cytometry

利用样品进样环通过流式细胞术精确测量颗粒数浓度

阅读:1

Abstract

Flow cytometry plays a pivotal role in biotechnology by providing quantitative measurements for a wide range of applications. Nonetheless, achieving precise particle quantification, particularly without relying on counting beads, remains a challenge. In this study, we introduce a novel exhaustive counting method featuring a sample loop-based injection system that delivers a defined sample volume to a detection system to enhance quantification in flow cytometry. We systematically assess the performance characteristics of this system with micron-sized polystyrene beads, addressing issues related to sample introduction, adsorption, and volume measurement. Results underscore the excellent analytical performance of the proposed method, characterized by high linearity and repeatability. We compare our approach to counting bead-based measurements, and while an approximate bias value was observed, the measured values were found to be similar between the methods, demonstrating its comparability and reliability. This method holds great promise for improving the accuracy and precision of particle quantification in flow cytometry, with implications for various fields including healthcare and environmental monitoring.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。