Validation of Macular Choroidal Thickness Measurements from Automated SD-OCT Image Segmentation

基于自动SD-OCT图像分割的黄斑脉络膜厚度测量的验证

阅读:1

Abstract

PURPOSE: Spectral domain optical coherence tomography (SD-OCT) imaging permits in vivo visualization of the choroid with micron-level resolution over wide areas and is of interest for studies of ocular growth and myopia control. We evaluated the speed, repeatability, and accuracy of a new image segmentation method to quantify choroid thickness compared to manual segmentation. METHODS: Two macular volumetric scans (25 × 30°) were taken from 30 eyes of 30 young adult subjects in two sessions, 1 hour apart. A single rater manually delineated choroid thickness as the distance between Bruch's membrane and sclera across three B-scans (foveal, inferior, and superior-most scan locations). Manual segmentation was compared to an automated method based on graph theory, dynamic programming, and wavelet-based texture analysis. Segmentation performance comparisons included processing speed, choroid thickness measurements across the foveal horizontal midline, and measurement repeatability (95% limits of agreement (LoA)). RESULTS: Subjects were healthy young adults (n = 30; 24 ± 2 years; mean ± SD; 63% female) with spherical equivalent refractive error of -3.46 ± 2.69D (range: +2.62 to -8.50D). Manual segmentation took 200 times longer than automated segmentation (780 vs. 4 seconds). Mean choroid thickness at the foveal center was 263 ± 24 μm (manual) and 259 ± 23 μm (automated), and this difference was not significant (p = 0.10). Regional segmentation errors across the foveal horizontal midline (±15°) were ≤9 μm (median) except for nasal-most regions closest to the nasal peripapillary margin-15 degrees (19 μm) and 12 degrees (16 μm) from the foveal center. Repeatability of choroidal thickness measurements had similar repeatability between segmentation methods (manual LoA: ±15 μm; automated LoA: ±14 μm). CONCLUSIONS: Automated segmentation of SD-OCT data by graph theory and dynamic programming is a fast, accurate, and reliable method to delineate the choroid. This approach will facilitate longitudinal studies evaluating changes in choroid thickness in response to novel optical corrections and in ocular disease.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。