The measurement of entrance surface dose using optically stimulated luminescence dosimeters for determining average glandular dose in digital breast tomosynthesis: Measurement and simulation study

利用光刺激发光剂量计测量入射表面剂量以确定数字乳腺断层合成中平均腺体剂量:测量和模拟研究

阅读:1

Abstract

This study aimed to evaluate the feasibility of using optically stimulated luminescence dosimeters (OSLDs), nanoDots, for the determination of an average glandular dose (AGD) with a specific digital breast tomosynthesis (DBT) system, whereas the X-ray tube was fixed (2D mode) and moved (3D mode). The entrance surface air kerma (ESAK) was measured by placing the nanoDots on the surface of a polymethyl methacrylate (PMMA) phantom with 25, 28, and 34 kV W/Rh techniques. The experimental setup of the ESAK measurement was simulated using a Monte Carlo simulation code to determine the ESAK and the backscatter factor (BSF). The AGD was calculated by dividing the ESAK values over the corresponding BSF factors for each PMMA phantom thickness and multiplying the AGD conversion factors. The AGD determination by the nanoDots variated within ±5% for both 2D and 3D modes, compared to those determined using an ionization chamber. The results were similarly observed for the simulation, except for the 25 kV on the 3D mode. Regarding the International Atomic Energy Agency technical reports series number 457, the nanoDots can be used for the AGD determination with realistic 2D and 3D image acquisitions based on ±10% uncertainty.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。