Entrance Skin Dose (ESD) and Bucky Table Induced Backscattered Dose (BTI-BSD) in Abdominal Radiography With nanoDot Optically Stimulated Luminescence Dosimeter (OSLD)

利用纳米点光刺激发光剂量计 (OSLD) 测量腹部放射成像中的入射皮肤剂量 (ESD) 和巴基台诱导反散射剂量 (BTI-BSD)

阅读:1

Abstract

In radiography, inconsistencies in patients' measured entrance skin dose (ESD) exist. There is no published research on the bucky table induced backscattered radiation dose (BTI-BSD). Thus, we aimed to ascertain ESD, calculate the BTI-BSD in abdominal radiography with a nanoDot OSLD, and compare the ESD results with the published data. A Kyoto Kagaku PBU-50 phantom (Kyoto, Japan) in an antero-posterior supine position was exposed, selecting a protocol used for abdominal radiography. The central ray of x-ray beam was pointed at the surface of abdomen at the navel, where a nanoDot dosimeter was placed to measure ESD. For the BTI-BSD, exit dose (ED) was determined by placing a second dosimeter on the exact opposite side (backside) of the phantom from the dosimeter used to determine (ESD) with and without bucky table at identical exposure parameters. The BTI-BSD was calculated as the difference between ED with and without bucky table. The ESD, ED, and BTI-BSD were measured in milligray (mGy). ESD mean values with and without bucky table were 1.97 mGy and 1.84 mGy, whereas ED values were 0.062 mGy and 0.052 mGy, respectively. Results show 2-26% lower ESD values with nanoDot OSLD. The BTI-BSD mean value was found to be approximately 0.01 mGy. A local dose reference level (LDRL) can be established using ESD data to safeguard patients from unnecessary radiation. In addition, to minimize the risk of BTI-BSD in patients in radiography, the search for the use or fabrication of a new, lower atomic number material for the bucky table is suggested.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。