Influence of Physicochemical Properties of Budesonide Micro-Suspensions on Their Expected Lung Delivery Using a Vibrating Mesh Nebulizer

布地奈德微混悬液的理化性质对其使用振动网式雾化器进行肺部给药的预期效果的影响

阅读:1

Abstract

The efficiency of lung drug delivery of nebulized drugs is governed by aerosol quality, which depends both on the aerosolization process itself but also on the properties of aerosol precursors. This paper determines physicochemical properties of four analogous micro-suspensions of a micronized steroid (budesonide, BUD) and seeks relationships between these properties and the quality of the aerosol emitted from a vibrating mesh nebulizer (VMN). Despite the same BUD content in all tested pharmaceutical products, their physicochemical characteristics (liquid surface tension, viscosity, electric conductivity, BUD crystal size, suspension stability, etc.) are not identical. The differences have a weak influence on droplet size distribution in the mists emitted from the VMN and on theoretical (calculated) regional aerosol deposition in the respiratory system but, simultaneously, there is an influence on the amount of BUD converted by the nebulizer to aerosol available for inhalation. It is demonstrated that the maximum inhaled BUD dose is below 80-90% of the label dose, depending on the nebulized formulation. It shows that nebulization of BUD suspensions in VMN is sensitive to minor dissimilarities among analogous (generic) pharmaceutics. The potential clinical relevance of these findings is discussed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。