Development and Validation of an Improved HPLC-MS/MS Method for Quantifying Total and Unbound Lenalidomide in Human Plasma

开发和验证一种改进的HPLC-MS/MS方法,用于定量人血浆中总来那度胺和游离来那度胺

阅读:1

Abstract

BACKGROUND/OBJECTIVES: This study aimed to develop a fully validated HPLC-MS/MS method for quantifying total and unbound lenalidomide concentrations in human plasma. METHODS: Unbound concentrations were measured using plasma ultrafiltrate prepared with Amicon(®) Centrifugal Filters. Lenalidomide and lenalidomide-d5 (internal standard) were extracted from 50 μL of human plasma using liquid-liquid extraction. Chromatography was conducted with a Halo(®) C18 column using 0.1% formic acid and methanol (20:80, v/v) as the mobile phase. The mass spectrometer was operated in a positive ion mode with an electrospray ionization interface and multiple reaction monitoring modes. RESULTS: Calibration curves were linear over the range of 5 to 1000 ng/mL (r(2) > 0.996) for both the total and unbound lenalidomide. For total lenalidomide concentrations, between-run precision (coefficients of variation) and accuracy were 1.70-7.65% and 94.45-101.10%, respectively. For unbound concentrations, inter-day precision and accuracy were 1.98-10.55% and 93.95-98.48%, respectively. CONCLUSIONS: We developed a highly reproducible, sensitive, and efficient bioanalytical method using a smaller volume of plasma sample (50 μL) with a relatively short run time (2.5 min). The proposed analytical method was successfully applied to measure total and unbound lenalidomide concentrations at various time points in multiple myeloma patients with renal impairment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。